Column conditioning:
The general procedure of column conditioning is as follows:
1. Wash the column with about 3 column volumes of aqueous mobile Phase with a low flow rate.
2. Run linear gradient system from 100% aqueous mobile phase to 100% organic mobile phase up to 2-3 column volume at the identical flow rate as above.
3. Equilibrate the column on a minimum of 5 column volumes with the organic mobile phase, until the monitor signals get stable.
4. If you change the mobile phase system, a blank run should be done to verify any impurity that may emerge in the mobile phase.
Column cleaning:
Regularly cleaning the HPLC column is essential, particularly when dealing with samples that have any column foul contaminants and particulate matter. Increased back pressures, fluctuating baseline, and loss of resolution are all signs that the column may require to be cleaned. In silica-based reversed-phase media, the peak has loss resolution because of peak branding and it may also be caused by dissolution of the matrix during regular use of mobile phase or buffer of above pH 7.
A typical procedure for cleaning an HPLC column is as follows.
- All the solvent using in the analysis should be HPLC grade and filtered through the 0.45-micron nylon filter membrane and degassing is also necessary.
- Equilibrate the column at a low flow rate with several column volumes of mobile phase containing HPLC grade water.
- 100% - 20 column volumes of water
- 50:50% - 20 column volumes of water: acetonitrile or methanol
- 100% - 20 column volumes of acetonitrile or methanol.The buffer solutions are very harmful to the performance and lifespan of the column, hence wash the column with an appropriate volume of water, so that salt does not get accumulated in the column and system.
Column storage:
Regeneration of the reversed-phase column:
The RP-HPLC columns include C8, C18, CN C1, C30, C4, and phenyl stationary phases. To wash and generate the column, flush it with the following solvents.
- 20 column volumes of 95:5% of water: acetonitrile (v/v)
- 20 column volumes of 100 % of acetonitrile
- 5 column volumes 100 % of isopropanol
- 20 column volumes 100 % of hexane
- 5 column volumes 100 % of isopropanol
- 20 column volumes 100 % of acetonitrile
- 20 column volumes of water: acetonitrile (95:5 v/v)
Important points:
- Properly install the column when you change or replace it.
- Confirm the efficiency of the new column supplied by the vendor.
- The column should be protected from knocks, bumps, and mechanical shock.
- To protect against impurities and other foreign particulates, always use a guard column.
- Always use HPLC grade solvents (methanol, Acetonitrile, and water, etc.) for analysis.
- To exclude impurities and foreign particles, filter out the mobile phase using 0 .45 μ and before injecting the sample filter the sample using a 0 .20 μ membrane filter.
- Degas all mobile phases and samples before use for the analysis.
- Always use freshly prepared buffer solution (phosphate, acetate, and formate, etc.) for analysis.
- Before injecting the sample, saturate the column with the mobile phase.
- Do not overload the HPLC column resulting in broad peaks of analytes.
- If a column oven is available then maintain column temperature, and also maintain the room temperature.
- Always maintain the pH of the buffer/mobile phase in the range of 2.0 to 9.5 pH.
- Avoid using a viscous buffer. Phosphate buffer concentrations should be less than 20 mM.
- Avoid injecting the highly acidic and basic sample into the sample loop, always neutralize the sample.
- Do not inject biological samples directly into the HPLC injector.
- Avoid any sudden pressure fluctuation during the analysis.
- Do not work with high flow rates, it creates pressure
- Periodically check the performance of the column with the standards and do not operate HPLC system at back pressure greater than 4000 psi.
- After each use, thoroughly clean the column with HPLC grade water and a suitable solvent.
- Avoid storing the column with buffers, store it with the necessary solvents (methanol/acetonitrile), and do not allow it to dry.
You may also like this
HPLC Columns
HPLC column care and maintenance
Retention Time
Factor Affecting Retention Time in HPLC
Resolution in Chromatography
Dead Volume
HPLC Column Regeneration
Role of ion-pairing reagents in HPLC
Sample Injection System in HPLC
Selection of buffer in HPLC method development
What is an ion-pairing reagent and why is used in HPLC
What are the common buffers used in HPLC?
No comments:
Post a Comment